# Fluke 110 multimeter: specs

> Handheld 6000-count true-RMS digital multimeter without a current input. Rated CAT III 600 V by Fluke.

[HTML version](https://soldr.ai/docs/components/fluke-110)

[Docs](https://soldr.ai/docs) / [Components](https://soldr.ai/docs/components)

Handheld 6000-count true-RMS digital multimeter without a current input. Rated CAT III 600 V by Fluke.

Every value cites its sourcePinout not published

[Buy on Compoden](https://compoden.com/products/fluke-110-true-rms-digital-multimeter)

## Specs

| Spec | Value | Source |
|---|---|---|
| DC voltage | 600.0 mV to 600.0 V; accuracy 0.5 % + 2 | source Specifications, Table 6 Accuracy (p.16) |
| AC voltage (true-RMS) | 600.0 mV to 600.0 V; 1.0 % + 3 (45 Hz to 500 Hz), 2.0 % + 3 (500 Hz to 1 kHz) | source Specifications, Table 6 Accuracy (p.16) |
| Resistance | 600.0 ohm to 40.00 Mohm; 0.9 % + 2 (600.0 ohm), 0.9 % + 1 (6.000 kohm to 6.000 Mohm), 5.0 % + 2 (40.00 Mohm) | source Table 6 Accuracy Specifications (p.16), model column 110 |
| Continuity | 600 ohm range; beeper on below 20 ohm, off above 250 ohm; detects opens or shorts of 500 us or longer | source Table 6 Accuracy Specifications (p.16), model column 110 |
| Current input | None: the amps input terminal is listed for the 115 and 117 only | source Table 3 Terminals (p.4) |
| Safety | IEC 61010-2-033 Measurement CAT III 600 V; 600 V maximum between any terminal and earth ground | source Specifications (p.14-15) |
| Display | 6000 counts, updates 4 per second | source Specifications (p.14) |
| Battery | 9 V alkaline IEC 6LR61; 400 hours typical without backlight | source Specifications (p.15) |
| Operating temperature | -10 C to 50 C (storage -40 C to 60 C) | source Specifications (p.15) |
| Ingress protection | IP42 (non-operating) | source Specifications (p.15) |

## Power

| | Value | Source |
|---|---|---|
| Supply | 9 V alkaline battery (IEC 6LR61) | source Specifications (p.15) |
| Current | 400 hours typical without backlight | source Specifications (p.15) |

## Pinout
We don't publish a pin table for this part: no official source we could verify gives one. Check the labels printed on your module, and see the notes below.

A meter has no header; see the Fluke user manual (110/113/114/115/117) for jack positions.

## Build with it
Project ideas for the Fluke 110 multimeter. Each link opens Soldr with the prompt filled in; Soldr assigns the pins, checks power and shows how to wire every part.

- [Battery pack voltage under load](https://app.soldr.ai/?prompt=I%20want%20to%20build%20a%20battery%20pack%20health%20check%20measuring%20a%209%20V%20six-cell%20AA%20pack%20with%20a%20Fluke%20110%20while%20it%20powers%20an%20Arduino%20Uno%20and%20an%20LCD): DC volts only on a 9 V pack, well inside the 600 V rating. Compare readings with and without the load.
- [A 5 V regulator output check with a Fluke 110 measuring a 7805 regulator fed from a 9 V six-cell AA pack](https://app.soldr.ai/?prompt=I%20want%20to%20build%20a%205%20V%20regulator%20output%20check%20with%20a%20Fluke%20110%20measuring%20a%207805%20regulator%20fed%20from%20a%209%20V%20six-cell%20AA%20pack): A 0.5 % + 2 DC volts reading shows whether the output is inside the regulator&#x27;s 4.8 V to 5.2 V range. Keep the load small so the 7805 stays cool.
- [Set a buck converter to 5.0 V](https://app.soldr.ai/?prompt=I%20want%20to%20build%20a%205.0%20V%20setup%20check%20where%20a%20Fluke%20110%20verifies%20a%20buck%20converter%20module%20before%20it%20powers%20an%20ESP32%20board): Adjust the converter with no load, read it with the meter, then connect the ESP32 board. This protects the 5 V input of the board.

## Common mistakes

- Looking for a current jack: the 110 has no amps input, so it cannot measure circuit current ([source](https://media.fluke.com/5e1db354-3a6e-49cf-9edd-b10800c0e608_original%20file.pdf))
- Using it above its rating: 600 V to earth ground and CAT III 600 V; mains wiring needs a qualified person ([source](https://media.fluke.com/5e1db354-3a6e-49cf-9edd-b10800c0e608_original%20file.pdf))
- Reading outside the rated conditions: accuracy is specified at 18 C to 28 C ([source](https://media.fluke.com/5e1db354-3a6e-49cf-9edd-b10800c0e608_original%20file.pdf))

## Not covered by the official source

- The 110 has no capacitance or frequency function: Table 6 lists capacitance for the 113, 115 and 117, and frequency for the 115 and 117.
- All values are from the Fluke 110/113/114/115/117 users manual (Table 6 and the general specifications); only rows whose model column includes the 110 are used.